Abstract

Wearable devices for digital continuous glucose monitoring (CGM) have attracted great attention as a new paradigm medical device for diabetes management. However, the relatively inaccurate performance and instability of CGM devices have limited their wide applications in the clinic. Here, we developed hyaluronate (HA) modified Au@Pt bimetallic electrodes for long-term accurate and robust CGM of smart contact lens. After glucose oxidation reaction, the bimetallic electrodes facilitated the rapid decomposition of hydrogen peroxide and charge transfer for robust CGM. The passivation of Au@Pt bimetallic electrode with branch-type thiolated HA prevented the dissolution of Au electrode by chloride ions in tears. In diabetic and normal rabbits, the smart contact lens with HA-Au@Pt bimetallic electrodes enabled the high correlation (ρ = 0.88) CGM with 98.6% clinically acceptable data for 3 weeks. Taken together, we could confirm the feasibility of our smart contact lens for long-term CGM for further clinical development. [Display omitted] • The remarkable sensitivity and stability of glucose sensor by bimetallic HA-Au@Pt electrode. • The successful integration of a glucose sensor, a chip, and a wireless antenna in a contact lens. • A long-term stable wireless smart contact lens for robust digital diabetes diagnosis. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01429612
Volume :
302
Database :
Academic Search Index
Journal :
Biomaterials
Publication Type :
Academic Journal
Accession number :
173342703
Full Text :
https://doi.org/10.1016/j.biomaterials.2023.122315