Standardization of Rn-222 concentration using the multi-electrode proportional counter.
- Source :
-
Applied Radiation & Isotopes . Dec2023, Vol. 202, pN.PAG-N.PAG. 1p. - Publication Year :
- 2023
-
Abstract
- Standardization of the concentration of gaseous 222Rn based on a multi-electrode proportional counter (MEPC) is under development as a primary standard in Japan. In this study, the concept and evaluation of its performance are reported. The latter consists of a preliminary result for the uncertainty budget associated with the measurement of the MEPC and compensation of the electric field distortion in the MEPC. Moreover, an ionization-chamber-based gas circulation system was added for the calibration of radon monitors in the air. • Highlights paper 70 Furukawa et al.: • Standardization of the concentration of gaseous 222Rn based on a multi-electrode proportional counter (MEPC) is under development. • The relative standard uncertainty associated with the measurement of the MEPC was evaluated preliminarily to be 1.4%. • A new calibration system was devised that includes an inner-through-type ionization chamber as a transfer standard. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 09698043
- Volume :
- 202
- Database :
- Academic Search Index
- Journal :
- Applied Radiation & Isotopes
- Publication Type :
- Academic Journal
- Accession number :
- 173314315
- Full Text :
- https://doi.org/10.1016/j.apradiso.2023.111076