Back to Search Start Over

Implementation of BFP-Typed Apertometer: Reference Location and Accuracy.

Authors :
Zhang, Bei
Zhang, Cuiling
Source :
IEEE Transactions on Instrumentation & Measurement. 2021, Vol. 70, p1-8. 8p.
Publication Year :
2021

Abstract

This work is to investigate the instrumentation implementing issues of back focal plane (BFP)-typed apertometer. Firstly, we introduce a general instrument architecture of the BFP-typed apertometer, including the hardware (optics and mechanics) of the system, operation procedures of control and measurement system, and also the digital image processing algorithm/software. Secondly, we are the first to discuss the gradual rising effect of total-internal-reflection (TIR) phenomenon, the criteria on how to “exactly” locate the reference ring on BFP images, and the corresponding accuracies. As a significantly important complement, we further discuss three important issues: 1) the influence of Goos–Hänchen shift on numerical aperture (NA) measurement; 2) comparison of one-dimension and two-dimension identification algorithms; and 3) comparison of center locating using Fourier correlation analysis (FCA) and Hough transformation (HT) algorithms. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00189456
Volume :
70
Database :
Academic Search Index
Journal :
IEEE Transactions on Instrumentation & Measurement
Publication Type :
Academic Journal
Accession number :
170415505
Full Text :
https://doi.org/10.1109/TIM.2021.3077990