Uncertainty Evaluation of a Broadband Attenuation Standard.
- Source :
-
IEEE Transactions on Instrumentation & Measurement . Apr2005, Vol. 54 Issue 2, p705-708. 4p. - Publication Year :
- 2005
-
Abstract
- A standard step attenuator has been developed. It comprises a build-up chain of ten steps of power ratio measurement and operates in a frequency range of 10 MHz to 18 GHz. The overall system has a dynamic range of 90 dB and an expanded uncertainty of 0.006 dB (k = 2). The system is used as an attenuation standard in a series radio-frequency substitution scheme. The uncertainty factors in the attenuation measurement of 40 dB at S GHz are fully described. For the calibration of a typical step attenuator, the expanded uncertainties are 0.011 dB for the attenuation range up to 80 dB, including the mismatch uncertainty. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00189456
- Volume :
- 54
- Issue :
- 2
- Database :
- Academic Search Index
- Journal :
- IEEE Transactions on Instrumentation & Measurement
- Publication Type :
- Academic Journal
- Accession number :
- 16592292
- Full Text :
- https://doi.org/10.1109/TIM.2004.843336