Back to Search Start Over

Uncertainty Evaluation of a Broadband Attenuation Standard.

Authors :
Joo-Gwang Lee
Jeong-Hwan Kim
Jeong-Il Park
Ung-Taeg Kang
Source :
IEEE Transactions on Instrumentation & Measurement. Apr2005, Vol. 54 Issue 2, p705-708. 4p.
Publication Year :
2005

Abstract

A standard step attenuator has been developed. It comprises a build-up chain of ten steps of power ratio measurement and operates in a frequency range of 10 MHz to 18 GHz. The overall system has a dynamic range of 90 dB and an expanded uncertainty of 0.006 dB (k = 2). The system is used as an attenuation standard in a series radio-frequency substitution scheme. The uncertainty factors in the attenuation measurement of 40 dB at S GHz are fully described. For the calibration of a typical step attenuator, the expanded uncertainties are 0.011 dB for the attenuation range up to 80 dB, including the mismatch uncertainty. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00189456
Volume :
54
Issue :
2
Database :
Academic Search Index
Journal :
IEEE Transactions on Instrumentation & Measurement
Publication Type :
Academic Journal
Accession number :
16592292
Full Text :
https://doi.org/10.1109/TIM.2004.843336