Back to Search Start Over

Comparison of Javal-Schiøtz keratometer, Orbscan IIz and Pentacam topographers in evaluating anterior corneal topography.

Authors :
Turner, Jennifer M
Purslow, Christine
Murphy, Paul J
Source :
Clinical & Experimental Optometry. Jul2023, Vol. 106 Issue 5, p476-483. 8p.
Publication Year :
2023

Abstract

Inter-instrument variation in anterior corneal shape (ACS) measurement has a consequence for ocular clinical practice. To consider inter-instrument variability in keratometry measurements across the ACS and to explore instrument protocols for determining ACS keratometric analogues (KAs). Mean keratometry/KAs of the right eye were recorded using Javal-Schiøtz keratometer (J-S), Orbscan IIz and Pentacam from 124 subjects (78 females; mean ± SD age: 24.71 ± 6.61 years). Mean radii of curvature were obtained for 1-mm wide annular zones extending up to 6 mm (horizontally) and 4 mm (vertically) from the apex for Orbscan and Pentacam. Zonal mean radius of curvature was calculated by averaging keratometry values for all measured points within the zone. KA (mean ± SD): Horizontal: Orbscan (7.80 ± 0.31 mm) and J-S (7.82 ± 0.29 mm) were not significantly different (p = 0.072). Pentacam (7.86 ± 0.29 mm) was significantly flatter than J-S (p < 0.001) and Orbscan (p < 0.001). Vertical: Orbscan (7.64 ± 0.31 mm) was significantly steeper than J-S (7.67 ± 0.29 mm, p < 0.005) and Pentacam (7.70 ± 0.29 mm, p < 0.001). Pentacam was significantly flatter than J-S (p < 0.001) and significant flatter than Orbscan across the entire profile (1-4 mm zones horizontal and vertical, p < 0.001). LoAs (CI): J-S/Orbscan: ±0.75 mm (0.05–0.18); J-S/Pentacam: ±0.72 mm (0.01–0.12); Pentacam/Orbscan: ±0.16 mm (0.04–0.08). There was a +0.03 mm positive bias for Orbscan compared to J-S, +0.06 mm positive bias for Orbscan compared to Pentacam and −0.03 mm negative bias for Pentacam compared to J-S. Algorithms used by Orbscan and Pentacam to solve the peripheral paraxial ray problem produce significantly different KAs. Instrument-specific KAs cannot be used inter-changeably between instruments. Differences in KA between instruments are not significant for ocular surgery, but may influence rigid contact lens fitting. Pentacam measures flatter than Orbscan and J-S. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
08164622
Volume :
106
Issue :
5
Database :
Academic Search Index
Journal :
Clinical & Experimental Optometry
Publication Type :
Academic Journal
Accession number :
164784455
Full Text :
https://doi.org/10.1080/08164622.2022.2067470