Back to Search Start Over

CH4/C2H6 dual gas sensing system using a single mid-infrared laser.

Authors :
Sun, Jiachen
Chang, Jun
Zhang, Yu
Wei, Yubin
Zhang, Qinduan
Wang, Fupeng
Lin, Shan
Wang, Zihan
Mao, Minghui
Source :
Spectrochimica Acta Part A: Molecular & Biomolecular Spectroscopy. Apr2023, Vol. 291, pN.PAG-N.PAG. 1p.
Publication Year :
2023

Abstract

[Display omitted] • The CH 4 /C 2 H 6 dual gas sensor using a single mid-infrared laser is developed. • The cross-interference of C 2 H 6 on CH 4 is eliminated. • High sensitivity, strong stability, fast response speed and simple system structure. Methane (CH 4) and ethane (C 2 H 6) dual gas sensor with low system complexity and strong stability is proposed. The correction method based on absorbance spectrum is applied, and the cross-interference of C 2 H 6 to CH 4 is eliminated. In the single gas concentration measurement, linear fitting is performed between the absorbance and concentration of CH 4 and C 2 H 6 , and the correlation coefficients of R2 = 0.99959 and R2 = 0.99994 are obtained respectively, which proves that the accuracy of the dual gas sensor is robust. In the dual gas concentration measurement, we carry out continuous measurement of five mixed gases and a long-term measurement of a mixture of gases, which verifies that our sensor has the fast response speed and strong stability. The minimum detectable column densities of 0.62 ppm·m for CH 4 and 0.1 ppm·m for C 2 H 6 are achieved, respectively. The CH 4 /C 2 H 6 dual gas sensor assisted by the correction method has high sensitivity and strong robustness to cross-interference, and has great potential for application in various scenarios. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
13861425
Volume :
291
Database :
Academic Search Index
Journal :
Spectrochimica Acta Part A: Molecular & Biomolecular Spectroscopy
Publication Type :
Academic Journal
Accession number :
161905630
Full Text :
https://doi.org/10.1016/j.saa.2023.122368