Non-invasive phakometric measurement of corneal and crystalline lens alignment in human eyes.

Authors :
Dunne, Mark C. M.
Davies, Leon N.
Mallen, Edward A. H.
Kirschkamp, Thomas
Barry, Jean-Cyriaque
Source :
Ophthalmic & Physiological Optics. Mar2005, Vol. 25 Issue 2, p143-152. 10p.
Publication Year :
2005

Abstract

We describe a non-invasive phakometric method for determining corneal axis rotation relative to the visual axis (β) together with crystalline lens axis tilt (α) and decentration (d) relative to the corneal axis. This does not require corneal contact A-scan ultrasonography for the measurement of intraocular surface separations. Theoretical inherent errors of the method, evaluated by ray tracing through schematic eyes incorporating the full range of human ocular component variations, were found to be larger than the measurement errors (β < 0.67°,α < 0.72° andd < 0.08 mm) observed in nine human eyes with known ocular component dimensions. Intersubject variations (mean ± S.D.:β = 6.2 ± 3.4° temporal,α = 0.2 ± 1.8° temporal andd = 0.1 ± 0.1 mm temporal) and repeatability (1.96 × S.D. of difference between repeat readings:β ± 2.0°,α ± 1.8° andd ± 0.2 mm) were studied by measuring the left eyes of 45 subjects (aged 18–42 years, 29 females and 16 males, 15 Caucasians, 29 Indian Asians, one African, refractive error range−7.25 to +1.25 D mean spherical equivalent) on two occasions. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
02755408
Volume :
25
Issue :
2
Database :
Academic Search Index
Journal :
Ophthalmic & Physiological Optics
Publication Type :
Academic Journal
Accession number :
16107996
Full Text :
https://doi.org/10.1111/j.1475-1313.2004.00267.x