Back to Search Start Over

High-precision copper isotopic analysis using a Nu Sapphire MC-ICP-MS.

Authors :
Wang, Jing
Su, Ben-Xun
Tang, Dong-Mei
Yuan, Qing-Han
Li, Wen-Jun
Gao, Bing-Yu
Bao, Zhi-An
Zhao, Ye
Source :
JAAS (Journal of Analytical Atomic Spectrometry). Dec2022, Vol. 37 Issue 12, p2589-2598. 10p.
Publication Year :
2022

Abstract

Here we present an analytical method for Cu isotopic measurement using standard-sample bracketing on a Nu Sapphire multi-collector inductively coupled plasma mass spectrometer. The effects of Cu concentration and HNO3 molarity mismatch between the standard and the sample, as well as the presence of matrix elements have been evaluated. Intermediate precision of better than ±0.030‰ for δ65Cu (2SD) was routinely obtained on NIST SRM 3114 and NWU-Cu-A. Accurate measurements were achieved when CNa/CCu < 0.5, CTi/CCu, CFe/CCu and CCo/CCu < 1, CNi/CCu < 3 and CZn/CCu < 10. High-precision Cu isotopic data were determined for nineteen geological reference materials, including igneous rocks, sedimentary rocks, metamorphic rocks and chalcopyrite, with an overall range of ∼0.44‰ (−0.056‰ to 0.384‰), nearly 15 times the current analytical precision. The Cu isotopic data of these reference materials can be used for future interlaboratory comparisons. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
02679477
Volume :
37
Issue :
12
Database :
Academic Search Index
Journal :
JAAS (Journal of Analytical Atomic Spectrometry)
Publication Type :
Academic Journal
Accession number :
160507815
Full Text :
https://doi.org/10.1039/d2ja00197g