Comparison of a 22G Crown-Cut Needle with a Conventional 22G Needle with EBUS Guidance in Diagnosis of Sarcoidosis.

Authors :
Wälscher, J.
Büscher, E.
Bonella, F.
Karpf-Wissel, R.
Costabel, U.
Theegarten, D.
Rawitzer, J.
Wienker, J.
Darwiche, K.
Source :
Lung. Oct2022, Vol. 200 Issue 5, p633-641. 9p.
Publication Year :
2022

Abstract

Introduction: Endobronchial ultrasound transbronchial needle aspiration (EBUS-TBNA) is a standard procedure in cases of enlarged mediastinal lymph nodes. Recently, new tools were developed aiming to improve the diagnostic yield. A novel crown-cut needle is considered to obtain tissue cores which can be beneficial for the evaluation by the pathologist. This study aimed to compare the novel 22G crown-cut needle with a conventional 22G needle with EBUS guidance in the diagnosis of sarcoidosis. Methods: We designed a single-center prospective randomized clinical trial between March 2020 and January 2021 with 30 patients with mediastinal lymphadenopathy and suspected sarcoidosis. Results: 24 patients (mean age 49.5 vs 54.1, mean FVC 73.7% vs 86.7%, mean DLCO 72.4% vs 72.5% for crown-cut needle vs conventional needle, respectively) were diagnosed with sarcoidosis. In the remaining six patients, sarcoidosis was reasonably excluded. The diagnostic yield for sarcoidosis was 77% with the crown-cut needle vs. 82% with the conventional needle (p > 0.05). In patients with histopathologic hallmarks typical of sarcoidosis (n = 19), the crown-cut needle was superior in detecting granulomas (8.3 vs 3.8 per cytoblock, p < 0.05) and histiocytes (502 vs 186 per cytoblock, p < 0.05). Four of seven bronchoscopists experienced difficulties passing through the bronchial wall with the crown-cut needle and one episode of bleeding occurred in this group which made interventions necessary. Conclusions: Despite equivalence in diagnostic accuracy, the crown-cut needle was superior to the conventional needle in detecting granulomas and histiocytes. This indicates greater potential for obtaining higher quality sample material with the crown-cut needle in cases of granulomatous inflammation. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
03412040
Volume :
200
Issue :
5
Database :
Academic Search Index
Journal :
Lung
Publication Type :
Academic Journal
Accession number :
159441461
Full Text :
https://doi.org/10.1007/s00408-022-00562-x