Back to Search Start Over

Demonstration of fast multiplicity counting of 252Cf using plastic scintillators and silicon photomultipliers.

Authors :
Cohen, E.O.
Waschitz, Y.
Ifergan, Y.
Roy, A.
Vartsky, D.
Yehuda-Zada, Y.
Israelashvili, I.
Beck, A.
Arazi, L.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Oct2022, Vol. 1040, pN.PAG-N.PAG. 1p.
Publication Year :
2022

Abstract

We present a first prototype for a multiplicity counter of fast neutrons and γ rays, based on plastic scintillators coupled to silicon photomultiplier arrays, where the SiPMs are read out individually by the TOFPET2 ASIC providing time-stamps and charge information. We demonstrate the capabilities of our counter by measuring the Singles-, Doubles- and Triples-event rates of neutrons and γ rays (without pulse shape discrimination), for a 252Cf source with a nominal activity of 4.9 μ Ci, and comparing them with a GEANT4 simulation of the process. [ABSTRACT FROM AUTHOR]