Early failure of an uncemented femoral stem, as compared to two other stems with similar design, following primary total hip arthroplasty performed with direct anterior approach.
- Source :
-
Hip International . Mar2022, Vol. 32 Issue 2, p166-173. 8p. - Publication Year :
- 2022
-
Abstract
- Introduction: An increased risk of early femoral component loosening has been reported using the direct anterior approach (DAA) compared with other common surgical approaches. However, long-term data are scarce. The purpose of this study is: (1) to determine the incidence of early femoral loosening in a high volume, single surgeon's practice utilising the DAA approach; and (2) to examine the effect of stem design and type of coating on aseptic loosening in the early and mid-term postoperative period. Methods: A retrospective review of 1650 consecutive patients (1800 hips) who underwent total hip arthroplasty (THA) using the DAA between August 2011 and December 2017 was conducted at our institution. 3 types of uncemented femoral stems (Quadra-S, Avenir, TwinSys), with similar design, but different coating, were implanted. Patients were evaluated clinically and radiologically at 4 weeks, 3 months, 1 year, and annually thereafter. Results: After a mean follow-up of 46.4 months, the total incidence of revision for aseptic loosening was 0.44% (n = 8). All loose stems were Quadra-S, failing to achieve osseointegration, for an overall incidence of 0.96% (p = 0.002). None of the other stems were loose. Radiolucent lines around the proximal stem portion were visible in 75 Quadra-S stems (4.1%) on radiographs taken at 1 year postoperatively and continued to deteriorate at the latest follow-up. Conclusions: In this specific cohort of patients, the increased rate of femoral stem aseptic loosening was implant-related and was attributed only to a specific type of femoral stem (Quadra-S). No relation to other factors was proven suggesting that the surface characteristics of this femoral stem and the lack of bioactive coating are responsible for the observed early femoral failures. These findings should be confirmed by additional registry work and larger population sample sizes are needed to evaluate the prosthesis performance after implantation through the DAA. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 11207000
- Volume :
- 32
- Issue :
- 2
- Database :
- Academic Search Index
- Journal :
- Hip International
- Publication Type :
- Academic Journal
- Accession number :
- 156076134
- Full Text :
- https://doi.org/10.1177/1120700020940671