Lowering the setting value of the esophageal endoscopic submucosal dissection device enabled tissue damage control in vitro porcine model.

Authors :
Yamaguchi, Yukiko
Uesato, Masaya
Yonemoto, Shohei
Maruyama, Tetsuro
Urahama, Ryuma
Suito, Hiroshi
Kishimoto, Takashi
Shiko, Yuki
Ozawa, Yoshihito
Kawasaki, Yohei
Matsubara, Hisahiro
Source :
Scientific Reports. 2/23/2022, Vol. 12 Issue 1, p1-7. 7p.
Publication Year :
2022

Abstract

One of the complications of esophageal endoscopic submucosal dissection (ESD) is postoperative stricture formation. Stenosis formation is associated with inflammation and fibrosis in the healing process. We hypothesized that the degree of thermal damage caused by the device is related to stricture formation. We aimed to reveal the relationship between thermal damage and setting value of the device. We energized a resected porcine esophagus using the ESD device (Flush Knife 1.5). We performed 10 energization points for 1 s, 3 s, and 5 s at four setting values of the device. We measured the amount of current flowing to the conducted points and the temperature and evaluated the effects of thermal damage pathologically. As results, the mean highest temperatures for 1 s were I (SWIFT Effect3 Wat20): 61.19 °C, II (SWIFT Effect3 Wat30): 77.28 °C, III (SWIFT Effect4 Wat20): 94.50 °C, and IV (SWIFT Effect4 Wat30): 94.29 °C. The mean heat denaturation areas were I: 0.84 mm2, II: 1.00 mm2, III: 1.91 mm2, and IV: 1.54 mm2. The mean highest temperature and mean heat denaturation area were significantly correlated (P < 0.001). In conclusion, Low-current ESD can suppress the actual temperature and thermal damage in the ESD wound. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20452322
Volume :
12
Issue :
1
Database :
Academic Search Index
Journal :
Scientific Reports
Publication Type :
Academic Journal
Accession number :
155397199
Full Text :
https://doi.org/10.1038/s41598-022-06533-9