In vivo human retinal swept source optical coherence tomography and angiography at 830 nm with a CMOS compatible photonic integrated circuit.
- Source :
-
Scientific Reports . 10/26/2021, Vol. 11 Issue 1, p1-12. 12p. - Publication Year :
- 2021
-
Abstract
- Photonic integrated circuits (PIC) provide promising functionalities to significantly reduce the size and costs of optical coherence tomography (OCT) systems. This paper presents an imaging platform operating at a center wavelength of 830 nm for ophthalmic application using PIC-based swept source OCT. An on-chip Mach–Zehnder interferometer (MZI) configuration, which comprises an input power splitter, polarization beam splitters in the sample and the reference arm, and a 50/50 coupler for signal interference represents the core element of the system with a footprint of only (12 × 5) mm 2 . The system achieves 94 dB imaging sensitivity with 750 μ W on the sample, 50 kHz imaging speed and 5.5 μ m axial resolution (in soft tissue). With this setup, in vivo human retinal imaging of healthy subjects was performed producing B-scans, three-dimensional renderings as well as OCT angiography. These promising results are significant prerequisites for further integration of optical and electronic building blocks on a single swept source-OCT PIC. [ABSTRACT FROM AUTHOR]