Back to Search Start Over

One‐year results of the first‐in‐man study investigating the Atrial Flow Regulator for left atrial shunting in symptomatic heart failure patients: the PRELIEVE study.

Authors :
Paitazoglou, Christina
Bergmann, Martin W.
Özdemir, Ramazan
Pfister, Roman
Bartunek, Jozef
Kilic, Teoman
Lauten, Alexander
Schmeisser, Alexander
Zoghi, Mehdi
Anker, Stefan D.
Sievert, Horst
Mahfoud, Felix
Source :
European Journal of Heart Failure. May2021, Vol. 23 Issue 5, p800-810. 11p. 3 Diagrams, 4 Charts, 1 Graph.
Publication Year :
2021

Abstract

Aims: Attenuating exercise‐induced elevated left atrial pressure with an atrial shunt device is under clinical investigation for treatment of symptomatic heart failure (HF). Methods and results: PRELIEVE was a prospective, non‐randomised, multicentre, first‐in‐man study in symptomatic HF patients with reduced (HFrEF) or preserved (HFpEF) ejection fraction and pulmonary capillary wedge pressure (PCWP) ≥15 mmHg at rest or ≥25 mmHg during exercise. Here, we provide follow‐up data up to 1 year after implantation of the Atrial Flow Regulator (AFR) device. The AFR was successfully implanted in 53 patients (HFrEF n = 24 and HFpEF n = 29). Two patients were not enrolled due to an unsuccessful transseptal puncture. There was one device embolisation into the left atrium, which required surgical removal. One patient experienced a serious procedure‐related adverse event (post‐procedural bleeding and syncope). All patients with sufficient echocardiography readout confirmed device patency with left–right shunt both at 3 (n = 47/51, 92%) and 12 (n = 45/49, 92%) months. At 3 months, rest PCWP decreased by 5 (−12, 0) mmHg (P = 0.0003, median Q1, Q3). No patient developed a stroke, worsening of right heart function or significant increase of pulmonary artery pressure. Six (6/53, 11%) patients were hospitalised for worsening of HF and three (3/53, 6%) patients died. We observed improvements in New York Heart Association functional class, 6‐min walking distance and quality of life (Kansas City Cardiomyopathy Questionnaire) in certain patients. Conclusions: Implantation of the AFR device in HF patients was feasible. No shunt occlusion, stroke or new right HF was observed during the 1‐year follow‐up, with clinical improvements in certain patients. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
13889842
Volume :
23
Issue :
5
Database :
Academic Search Index
Journal :
European Journal of Heart Failure
Publication Type :
Academic Journal
Accession number :
151135557
Full Text :
https://doi.org/10.1002/ejhf.2119