Actual delivered dose calculation on intra-irradiation cone-beam computed tomography images: a phantom study.

Authors :
Iramina, Hiraku
Nakamura, Mitsuhiro
Mizowaki, Takashi
Source :
Physics in Medicine & Biology. 1/7/2021, Vol. 66 Issue 1, p1-10. 10p.
Publication Year :
2021

Abstract

Cone-beam computed tomography (CBCT) images acquired during volumetric modulated arc therapy (VMAT; ii-CBCT) can be used to calculate actual delivered doses (ADDs). However, such ii-CBCT images are degraded by scattered megavoltage x-rays (MV-scatters). We aimed to evaluate the dose calculation accuracy of the MV-scatter uncorrected or corrected ii-CBCT images acquired during VMAT deliveries. For MV-scatter correction on concurrent kilovoltage projections (PMVkV), projections consisting only of MV-scatters (PMVS) were acquired under the same MV beam parameters and gantry angles and subtracted from PMVkV (PMVScorr). In addition, the projections by kilovoltage beams were acquired for reference (PkV). The corresponding CBCT images were reconstructed using the Feldkamp–Davis–Kress algorithm (CBCTMVkV, CBCTMVScorr, and CBCTkV as reference). A multi-energy phantom with rods of various relative electron densities (REDs) was used to generate a CBCT-number-to-RED conversion table. First, CBCTkV was reconstructed. Then, the mean CBCT-numbers within each rod were extracted, and a reference table was generated. Concurrent kilovoltage imaging with various field sizes was also demonstrated, and CBCTMVkV and CBCTMVScorr were reconstructed. The extracted CBCT-numbers of each ii-CBCT image were converted into REDs using the reference table. Next, the absolute differences of RED between the ii-CBCT image and CBCTkV were calculated. Ten VMAT plans using a 10 MV flattening-filter-free beam were used for concurrent imaging of an anthropomorphic torso phantom. Moreover, an iterative reconstruction algorithm (IRA) was used for CBCTMVScorr. The plans were recalculated for the corresponding CBCTMVkV, CBCTMVScorr, CBCTMVScorr+IRA, and CBCTkV with the reference table. Finally, the doses were evaluated using 3D gamma analysis (1%/1 mm). The median difference ranges between CBCTMVkV/CBCTMVScorr and the reference values were −0.58 to −0.10/−0.03 to 0.00. The median gamma pass rates of the doses on CBCTMVkV, CBCTMVScorr, and CBCTMVScorr+IRA to the rate on CBCTkV were 70.4, 99.5, and 98.2%, respectively. CBCTMVScorr were comparable with CBCTkV for calculating the ADD from VMAT. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00319155
Volume :
66
Issue :
1
Database :
Academic Search Index
Journal :
Physics in Medicine & Biology
Publication Type :
Academic Journal
Accession number :
149988084
Full Text :
https://doi.org/10.1088/1361-6560/abcdeb