Back to Search
Start Over
Development and Validation of an Automated Zone Fluidics-Based Sensor for In Vitro Dissolution Studies of Captopril Using Total Error Concept.
- Source :
-
Molecules . Feb2021, Vol. 26 Issue 4, p824-824. 1p. - Publication Year :
- 2021
-
Abstract
- In the present research, a zone fluidics-based automated sensor for the analysis of captopril in in vitro dissolution samples is reported. Captopril is reacted under flow conditions with Ni(II) (10 mmol L−1) in alkaline medium (0.15% v/v NH3) to form a stable derivate, which is monitored spectrophotometrically at 340 nm. The chemical and instrumental parameters were carefully investigated and optimized. The validation of the developed method was performed in the range of 5 to 120% of the expected maximum concentration using the accuracy profiles as a graphical decision-making tool. The β-expectation tolerance intervals did not exceed the acceptance criteria of ±10%, which means that 95% of future results will be encompassed in the defined bias limits. The variation of the relative bias ranged between −2.3% and 3.5% and the RSD values for repeatability and intermediate precision were lower than 2.3% in all cases. The limit of detection (LOD), and the lower and the upper limit of quantification (LLOQ, ULOQ) were satisfactory and found to be 1%, 5% and 120% (corresponding to 0.6, 2.78 and 66.67 μg mL−1 in dissolution medium). The developed method was successfully applied for the analysis of captopril in dissolution tests of two commercially available batches. [ABSTRACT FROM AUTHOR]
- Subjects :
- *CAPTOPRIL
*DETECTORS
*DETECTION limit
Subjects
Details
- Language :
- English
- ISSN :
- 14203049
- Volume :
- 26
- Issue :
- 4
- Database :
- Academic Search Index
- Journal :
- Molecules
- Publication Type :
- Academic Journal
- Accession number :
- 149852090
- Full Text :
- https://doi.org/10.3390/molecules26040824