Abstract
- We examine various options for calibration of the National Ignition Facility neutron detectors in the energy range E<14 MeV. These options include: downscatter of D–T fusion neutrons using plastic targets; nuclear reactions at a Tandem Van de Graaf accelerator; and “white” neutrons from a pulsed spallation source. As an example of the pulsed spallation option, we present calibration data that was recently acquired with a single crystal chemical vapor deposition diamond detector at the Weapons Neutron Research Facility at Los Alamos National Laboratory. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00346748
- Volume :
- 75
- Issue :
- 10
- Database :
- Academic Search Index
- Journal :
- Review of Scientific Instruments
- Publication Type :
- Academic Journal
- Accession number :
- 14842297
- Full Text :
- https://doi.org/10.1063/1.1779614