Comparison of Two Different Semiquantitative Urinary Dipstick Tests with Albumin-to-Creatinine Ratio for Screening and Classification of Albuminuria According to KDIGO. A Diagnostic Test Study.

Authors :
Hodel, Nikolai C.
Hamad, Ali
Reither, Klaus
Kasella, Irene Mndala
Abdulla, Salim
Schoetzau, Andreas
Hatz, Christoph F. R.
Mayr, Michael
Source :
Diagnostics (2075-4418). Jan2021, Vol. 11 Issue 1, p81-81. 1p.
Publication Year :
2021

Abstract

Background: Semiquantitative dipstick tests are utilized for albuminuria screening. Methods: In a prospective cross-sectional survey, we analyzed the diagnostic test validity of the semiquantitative colorimetric indicator-dye-based Combur9-Test® and the albumin-specific immunochromatographic assay Micral-Test® for the detection of albuminuria, the distribution of the semiquantitative measurements within the albuminuria stages according to KDIGO, and the utility for albuminuria screening compared with an albumin-to-creatinine ratio (ACR) in a walk-in population. Results: In 970 subjects, albuminuria (≥30 mg/g) was detected in 12.7% (95% CI 85.6–96.3%) with the ACR. Sensitivity was 82.9% (95% CI 75.1–89.1%) and 91.9% (95% CI 88.7–96.9%) and specificity 71.5% (95% CI 68.4–74.6%) and 17.5% (95% CI 15.0–20.2%) for the Combur9-Test® and Micral-Test®, respectively. Correct classification to KDIGO albuminuria stages A2/A3 with the Combur9-Test® was 15.4%, 51.4%, and 87.9% at cut-offs of 30, 100, and ≥300 mg/dL, and with the Micral-Test® it was 1.8%, 10.5%, and 53.6% at cut-offs of 2, 5, and 10 mg/dL, respectively. Overall, disagreement to KDIGO albuminuria was seen in 27% and 73% with the Combur9-Test® and Micral-Test®, respectively. From the total population, 62.5% and 15.3% were correctly ruled out and 2.2% and 1% were missed as false-negatives by the Combur9-Test® and Micral-Test®, respectively. Conclusion: Compared to the Combur9-Test®, the utility of the Micral-Test® is limited, because the fraction of correctly ruled out patients is small and a large proportion with a positive Micral-Test® require a subsequent ACR conformation test. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20754418
Volume :
11
Issue :
1
Database :
Academic Search Index
Journal :
Diagnostics (2075-4418)
Publication Type :
Academic Journal
Accession number :
148317583
Full Text :
https://doi.org/10.3390/diagnostics11010081