Back to Search Start Over

Porous Co3O4 modified carbon paste electrode for the quantification of dopamine.

Authors :
Shimna, Thooneri
Thomas, Jasmine
Thomas, Tony
Thomas, Nygil
Source :
Journal of Applied Electrochemistry. Nov2020, Vol. 50 Issue 11, p1165-1173. 9p.
Publication Year :
2020

Abstract

A simple Co3O4 modified carbon paste electrode (CPE) was developed as an electrochemical sensor for the selective detection and quantification of dopamine (DA) via fastest synthesis and fabrication methods. The solution combustion method and bulk modification techniques were sequentially used for the preparation of the sensor. The modifier was characterized using different techniques such as PXRD, FT-IR, SEM and TG-DSC. The fabricated sensor shows a wide linear dynamic range of 80 µM-300 nM, low detection limit, excellent reproducibility and high stability. The differential pulse voltammetry (DPV) and cyclic voltammetry (CV) techniques were extensively used for the electrochemical studies. The same electrode was used for detecting DA in the solution of coexisting ascorbic acid (AA) and uric acid (UA). Well resolved peaks and good current sensitivity of this electrochemical sensor extended its utility in determining DA in blood serum at physiological pH. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0021891X
Volume :
50
Issue :
11
Database :
Academic Search Index
Journal :
Journal of Applied Electrochemistry
Publication Type :
Academic Journal
Accession number :
146480193
Full Text :