Back to Search Start Over

Research on error correction method of filter assembly based on abnormal trend recognition of sequence data.

Authors :
Sun, Bojun
Sun, Xiaogang
Dai, Jingmin
Source :
Measurement (02632241). Oct2020, Vol. 163, pN.PAG-N.PAG. 1p.
Publication Year :
2020

Abstract

• Introduce a new method to identify the wrong filter assembly of pyrometer. • Reveal the relationship between radiation energy equivalence and wavelength. • The method is more effective for optical fiber type multi-wavelength pyrometers. Multi-wavelength pyrometer needs temperature calibration by blackbody before application. Aiming at multi-target multi-wavelength pyrometer with optical fiber for beam-splitting and filter for limiting wavelength, there are a large number of filters. Any error in the assembly of a filter will directly affect the accuracy of the measurement results. In this paper, an error correction method based on abnormal trend of temperature calibration sequence data is proposed for identifying whether the filter assembly is wrong or not. The intrinsic relationship between radiation energy equivalence and wavelength is revealed for the first time, providing theoretical basis for the method. And the simulation experiment and actual experiment are carried out to verify the feasibility of the method. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
02632241
Volume :
163
Database :
Academic Search Index
Journal :
Measurement (02632241)
Publication Type :
Academic Journal
Accession number :
145883637
Full Text :
https://doi.org/10.1016/j.measurement.2020.107932