Validation of an automated-ETDRS near and intermediate visual acuity measurement.

Authors :
Pang, Yi
Sparschu, Lauren
Nylin, Elyse
Source :
Clinical & Experimental Optometry. Sep2020, Vol. 103 Issue 5, p663-667. 5p.
Publication Year :
2020

Abstract

<bold>Background: </bold>The aim of this study was to determine the repeatability of an automated-ETDRS (Early Treatment Diabetic Retinopathy Study) near and intermediate visual acuity measurement in subjects with normal visual acuity and subjects with reduced visual acuity. The agreement of automated-ETDRS with gold standard chart-based visual acuity measurement was also studied.<bold>Methods: </bold>Fifty-one subjects were tested (aged 23 to 91 years; 33 subjects with normal visual acuity: 6/7.5 or better; 18 subjects with reduced visual acuity: 6/9 to 6/30). Near and intermediate visual acuity of one eye from each subject was measured with an automated tablet-computer system (M&S Technologies, Inc.) and Precision Vision paper chart in a random sequence. Subjects were retested one week later. Repeatability was evaluated using the 95 per cent limits of agreement (LoA) between the two visits.<bold>Results: </bold>Average difference between automated-ETDRS near visual acuity and near visual acuity by paper chart was 0.02 ± 0.10 logMAR (p > 0.05). Agreement of near visual acuity between automated-ETDRS and paper chart was good, with 95 per cent LoA of ±0.19 logMAR. Furthermore, automated-ETDRS near visual acuity showed good repeatability (95 per cent LoA of ±0.20). Mean difference between automated-ETDRS intermediate visual acuity and intermediate visual acuity by paper chart was 0.02 ± 0.10 logMAR (p > 0.05). Agreement of intermediate visual acuity between automated-ETDRS and paper chart was good, with 95 per cent LoA of ±0.20 logMAR. In addition, automated-ETDRS intermediate visual acuity had good repeatability (95 per cent LoA of ±0.16).<bold>Conclusion: </bold>Automated-ETDRS near and intermediate visual acuity measurement showed good repeatability and agreement with the gold standard chart-based visual acuity measurement. The findings of this study indicate the automated visual acuity measurement system may have potential for use in both patient care and clinical trials. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
08164622
Volume :
103
Issue :
5
Database :
Academic Search Index
Journal :
Clinical & Experimental Optometry
Publication Type :
Academic Journal
Accession number :
145298519
Full Text :
https://doi.org/10.1111/cxo.13018