Back to Search
Start Over
Challenges in Removing the Essure® Device.
- Source :
-
Case Reports in Obstetrics & Gynecology . 8/25/2020, p1-4. 4p. - Publication Year :
- 2020
-
Abstract
- We present a case about Essure® removal surgery in which the third markers of the device have torn off. The woman needed a second surgery for complete removal of the devices. Fluoroscopy during surgery is a good method to visualize the lost fragments. With fluoroscopy, a hysterectomy is not needed for complete removal. It is important to understand the structure of the device and to be aware of the four radiopaque markers during surgery and their removal. [ABSTRACT FROM AUTHOR]
- Subjects :
- *HYSTERECTOMY
*SURGERY
Subjects
Details
- Language :
- English
- ISSN :
- 20906684
- Database :
- Academic Search Index
- Journal :
- Case Reports in Obstetrics & Gynecology
- Publication Type :
- Academic Journal
- Accession number :
- 145298062
- Full Text :
- https://doi.org/10.1155/2020/8823606