Back to Search Start Over

Temperature Dependence of NaI(Tl) Radiation Scintillation Detectors' Characteristics.

Authors :
Csurgai, J.
Padányi, J.
Földi, L.
Source :
Advances in Military Technology. 2020, Vol. 15 Issue 1, p201-212. 12p.
Publication Year :
2020

Abstract

Scintillators belong to the oldest types of radiation detectors. Nowadays, for spectrometric purposes the use of semiconductive detectors is more common, but scintillation detectors are still in use for various specific measurement purposes. We have investigated the dependence of a gamma spectrum measured by NaI(Tl) scintillation detector on temperature changes. We examined the need of energy or efficiency recalibration and software compensation. Due to temperature dependence, scintillation detectors require energy recalibration before environmental and outdoor measurements or before the use of etalon sources for obtaining spectrum for follow-up calibration. We have elaborated on a simple method for energy recalibration of scintillation detectors at different temperatures. This method was converted into an algorithm and it can be either burned into instrument EPROM or used in software processing. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
18022308
Volume :
15
Issue :
1
Database :
Academic Search Index
Journal :
Advances in Military Technology
Publication Type :
Academic Journal
Accession number :
144595728
Full Text :
https://doi.org/10.3849/aimt.01328