Paper-Based Detection Device for Alzheimer's Disease—Detecting β-amyloid Peptides (1–42) in Human Plasma.

Authors :
Sung, Wei-Hsuan
Hung, Jung-Tung
Lu, Yu-Jen
Cheng, Chao-Min
Source :
Diagnostics (2075-4418). May2020, Vol. 10 Issue 5, p272. 1p.
Publication Year :
2020

Abstract

The diagnosis of Alzheimer's disease (AD) is frequently missed or delayed in clinical practice. To remedy this situation, we developed a screening, paper-based (P-ELISA) platform to detect β-amyloid peptide 1–42 (Aβ42) and provide rapid results using a small volume, easily accessible plasma sample instead of cerebrospinal fluid. The protocol outlined herein only requires 3 μL of sample per well and a short operating time (i.e., only 90 min). The detection limit of Aβ42 is 63.04 pg/mL in a buffer system. This P-ELISA-based approach can be used for early, preclinical stage AD screening, including screening for amnestic mild cognitive impairment (MCI) due to AD. It may also be used for treatment and stage monitoring purposes. The implementation of this approach may provide tremendous impact for an afflicted population and may well prompt additional and expanded efforts in both academic and commercial communities. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20754418
Volume :
10
Issue :
5
Database :
Academic Search Index
Journal :
Diagnostics (2075-4418)
Publication Type :
Academic Journal
Accession number :
144286693
Full Text :
https://doi.org/10.3390/diagnostics10050272