Back to Search Start Over

Solid-phase microextraction-capillary gas chromatography combined with microwave-induced plasma atomic-emission spectrometry for selenite determination.

Authors :
Dimitrakakis, Emmanouil
Haberhauer-Troyer, Christina
Abe, Yo
Ochsenkühn-Petropoulou, Maria
Rosenberg, Erwin
Source :
Analytical & Bioanalytical Chemistry. Jul2004, Vol. 379 Issue 5/6, p842-848. 7p.
Publication Year :
2004

Abstract

The use of solid-phase microextraction (SPME) with gas chromatography coupled to microwave-induced plasma atomic-emission detection (GC-MIP-AED) is described for selenite [Se(IV)] speciation. Aqueous standards were derivatised with sodium tetraethyl- or tetrapropylborate and extracted by SPME. Headspace extraction of the ethyl and propyl derivatives was studied. Relevant experimental conditions were optimised, including conditions for derivatisation and extraction and those of gas chromatographic analysis. The limits of detection achieved for headspace sampling of derivatised Se(IV) were in the low ng mL-1 range for both ethylation and propylation. When the method was applied to analysis of selenite in selenised yeast reference material results were in good agreement with the indicated values. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
16182642
Volume :
379
Issue :
5/6
Database :
Academic Search Index
Journal :
Analytical & Bioanalytical Chemistry
Publication Type :
Academic Journal
Accession number :
14387479
Full Text :