Back to Search Start Over

A multi-wavelength streaked optical pyrometer for dynamic shock compression measurements above 2500 K.

Authors :
Opachich, Y. P.
Crum, R. S.
Daene, M. W.
Dutra, E. C.
Mehta, H. N.
Shelton, H. L.
Watts, P.
Akin, M. C.
Source :
Review of Scientific Instruments. Mar2020, Vol. 91 Issue 3, p1-8. 8p. 5 Diagrams, 2 Charts, 5 Graphs.
Publication Year :
2020

Abstract

This article presents a concept and implementation of a calibrated streaked spectral pyrometer (SSP) temperature diagnostic used in dynamically driven shock experiments on a two-stage gas gun. This system relies upon measuring the total system response using a tunable monochromator, a NIST-traceable calibrated power meter, and a SSP. The diagnostic performance is validated against previously measured temperatures of shock driven z-cut quartz at 99 GPa and 93 GPa. The results are found to agree with the literature to within 5% and are discussed in this manuscript. The experimental setup utilizes measurements from multiple SSP systems per sample, providing several independent measurements and substantially increasing confidence in the extrapolated shock driven sample temperature. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00346748
Volume :
91
Issue :
3
Database :
Academic Search Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
142513927
Full Text :
https://doi.org/10.1063/1.5126123