A five‐axis parallel kinematic mirror unit for soft X‐ray beamlines at MAX IV.

Authors :
Agåker, Marcus
Mueller, Frieder
Norsk Jensen, Brian
Åhnberg, Karl
Sjöblom, Peter
Deiwiks, Jochen
Henniger, Hans
Pärna, Rainer
Knudsen, Jan
Thiagarajan, Balasubramanian
Såthe, Conny
Source :
Journal of Synchrotron Radiation. Mar2020, Vol. 27 Issue 2, p262-271. 10p.
Publication Year :
2020

Abstract

With the introduction of the multi‐bend achromats in the new fourth‐generation storage rings the emittance has decreased by an order of magnitude resulting in increased brightness. However, the higher brightness comes with smaller beam sizes and narrower radiation cones. As a consequence, the requirements on mechanical stability regarding the beamline components increases. Here an innovative five‐axis parallel kinematic mirror unit for use with soft X‐ray beamlines using off‐axis grazing‐incidence optics is presented. Using simulations and measurements from the HIPPIE beamline at the MAX IV Laboratory it is shown that it has no Eigen frequencies below 90 Hz. Its positioning accuracy is better than 25 nm linearly and 17–35 µrad angularly depending on the mirror chamber dimensions. [ABSTRACT FROM AUTHOR]