- Experiments performed at a temperature of a few millikelvins require effective thermalization schemes, low-pass filtering of the measurement lines, and low-noise electronics. Here, we report on the modifications to a commercial dilution refrigerator with a base temperature of 3.5 mK that enable us to lower the electron temperature to 6.7 mK measured from the Coulomb peak width of a quantum dot gate-defined in an [Al]GaAs heteostructure. We present the design and implementation of a liquid 4He immersion cell tight against superleaks, implement an innovative wiring technology, and develop optimized transport measurement procedures. [ABSTRACT FROM AUTHOR]