Back to Search Start Over

Medical, Technical and Audiological Outcomes of Hearing Rehabilitation with the Bonebridge Transcutaneous Bone-Conduction Implant: A Single-Center Experience.

Authors :
Brkic, Faris F.
Riss, Dominik
Scheuba, Katharina
Arnoldner, Christoph
Gstöttner, Wolfgang
Baumgartner, Wolf-Dieter
Vyskocil, Erich
Source :
Journal of Clinical Medicine. Oct2019, Vol. 8 Issue 10, p1614. 1p.
Publication Year :
2019

Abstract

Bone-conduction implants are a standard therapeutic option for patients with conductive, unilateral, or mixed hearing loss who either do not tolerate conventional hearing aids or can benefit from surgery. The aim of this study was to evaluate long-term medical and technical outcomes, and audiological results with the Bonebridge transcutaneous bone-conduction implant. This retrospective study included all patients implanted with a bone-conduction hearing implant at a tertiary medical referral center between March 2012 and October 2018. Medical and technical outcomes included the mean length of implant usage, medical and technical complications (skin and wound infection, lack of benefit, technical failure), explantations and revisions, coupling approaches, implant failure rate, implant survival and the implant loss for added follow-up years. Auditory results were measured by functional hearing gain and the Freiburger monosyllabic test at 65 dB sound pressure level. Sixty-four patients were included in the study; five of these were implanted bilaterally (69 devices). Five unilaterally implanted patients were lost to follow-up. The mean follow-up was 27.1 months (range: 0.2 months–6.3 years). The mean implant usage was 25.9 months (range: 0.2 months–6.3 years). Fifty-seven implants (89.1%) were in use at the end of the follow-up period. Complications occurred in six ears (9.4%). Five implants (7.8%) were explanted without reimplantation. Device failure occurred in one implant (1.6%), which was possibly caused by recurrent head trauma. The rate of implant loss due to technical device failure (damage to device) was 1 per 72 follow-up years. The mean improvement on the Freiburger monosyllabic test (52.1%, p = 0.0001), and in functional hearing gain across frequencies (26.5 dB, p = 0.0001) was significant. This single-center follow-up reveals the medical and technical reliability of a transcutaneous bone-conduction implant for hearing rehabilitation because complication and revision rates were low. The majority of patients still used the device at the end of the observation period. Implantation resulted in favorable hearing outcomes in comparison to that of unaided conditions. Cautious patient selection mainly regarding co-morbidities, the history of chronic otologic diseases and proper surgical technique seems to be crucial in reducing complications. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20770383
Volume :
8
Issue :
10
Database :
Academic Search Index
Journal :
Journal of Clinical Medicine
Publication Type :
Academic Journal
Accession number :
139330950
Full Text :
https://doi.org/10.3390/jcm8101614