Back to Search Start Over

Label-free electrochemical impedimetric immunosensor for sensitive detection of IgM rheumatoid factor in human serum.

Authors :
Chinnadayyala, Somasekhar R.
Park, Jinsoo
Abbasi, Muhammad A.
Cho, Sungbo
Source :
Biosensors & Bioelectronics. Oct2019, Vol. 143, pN.PAG-N.PAG. 1p.
Publication Year :
2019

Abstract

The authors report a label-free and direct detection of rheumatoid factor- Immunoglobulin M (IgM-RF) based on an impedimetric-interdigitated wave type microelectrode array (IDWμE). IDWμE was functionalized with a self-assembled monolayer (SAM) of thioctic acid for antigen immobilization. The SAM functionalized IDWμE were characterized by atomic force microscopy, Energy Dispersive X-Ray Spectroscopy, and X-ray photoelectron spectroscopy. The covalent immobilization of IgG-Fc onto the SAM modified electrode arrays was characterized morphologically via AFM and electrochemically via cyclic voltammetry and electrochemical impedance spectroscopy. Impedimetric measurements in the presence of redox probe (Potassium ferrocyanide and potassium ferricyanide) showed a significant change in both the impedance spectrum and charge transfer resistance upon IgM-RF binding. Impedance measurements were target specific and linear with an increase in IgM-RF concentrations between 1 and 200 IU mL−1 in redox probe and human serum, respectively. The sensor showed detection limits of 0.6 IU mL−1 in the presence of redox probe and 0.22 IU mL−1 in the human serum. The biosensor exhibited good reproducibility (relative standard deviation (RSD), 4.96%) and repeatability (RSD, 2.31%) with an acceptable selectivity towards IgM-RF detection. The sensor also showed a good stability for 3 weeks at 4 °C in 1X PBS. Therefore, the sensitive and stable immunosensor exhibiting IDWμE features can be integrated with a miniaturized potentiostat to develop a sensing system that detects IgM-RF for point-of-care applications. • IDWμE array was fabricated using photolithography and a chemical wet etching. • Immunosensor based on self-assembled monolayers of thioctic acid was prepared. • IgM-RF was detected using a sensitive label-free impedimetric Immunosensor. • Biosensor presents good stability and selectivity to quantify IgM-RF in serum samples. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09565663
Volume :
143
Database :
Academic Search Index
Journal :
Biosensors & Bioelectronics
Publication Type :
Academic Journal
Accession number :
138499856
Full Text :
https://doi.org/10.1016/j.bios.2019.111642