Back to Search Start Over

Reproducibility of broth microdilution MICs for the novel siderophore cephalosporin, cefiderocol, determined using iron-depleted cation-adjusted Mueller-Hinton broth.

Authors :
Hackel, Meredith A.
Tsuji, Masakatsu
Yamano, Yoshinori
Echols, Roger
Karlowsky, James A.
Sahm, Daniel F.
Source :
Diagnostic Microbiology & Infectious Disease. Aug2019, Vol. 94 Issue 4, p321-325. 5p.
Publication Year :
2019

Abstract

In 2017, the Clinical and Laboratory Standards Institute (CLSI) Subcommittee on Antimicrobial Susceptibility Testing approved the use of iron-depleted cation-adjusted Mueller-Hinton broth (ID-CAMHB) prepared with Chelex® 100 resin (Bio-Rad Laboratories, Hercules, CA) to determine MICs for cefiderocol. The current study examined the reproducibility of cefiderocol MICs generated for 19 clinical isolates of Gram-negative bacilli, with CAMHB produced by three manufacturers; each of the 19 isolates was tested for 10 replicates in ID-CAMHB from each manufacturer. When analyzed by individual media lot, greater than 95% of MIC results were within ± one doubling-dilution of the mode for each of the 19 isolates tested. The remaining 5.0% of MIC results were within ± two doubling-dilutions of the modal MIC. For all media lots combined, 92.2% of MIC results were within ± one doubling-dilution of the modal MIC for each isolate, 99.8% were within ± two doubling-dilutions and 100% were within three doubling-dilutions. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
07328893
Volume :
94
Issue :
4
Database :
Academic Search Index
Journal :
Diagnostic Microbiology & Infectious Disease
Publication Type :
Academic Journal
Accession number :
137361063
Full Text :
https://doi.org/10.1016/j.diagmicrobio.2019.03.003