Back to Search Start Over

Optimization of data acquisition parameters for single gamma line sensing: KSr2I5:Eu[formula omitted] and NaI(Tl) compared.

Authors :
Lukosi, Eric
Rust, Mikah
Stand, Luis
Melcher, Charles L.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Sep2019, Vol. 938, p36-40. 5p.
Publication Year :
2019

Abstract

KSr 2 I 5 :Eu 2 + scintillators exhibit better than 3% energy resolution at 662 keV and may be significantly less expensive to produce than other, similar energy resolution scintillators. However, KSr 2 I 5 :Eu 2 + exhibits a 6.5 Bq/cm3 intrinsic background from the presence of 40K. This background is not only the well-known 1.46 MeV gamma, but also the 1.31 MeV end point of the beta spectrum. This paper computationally evaluates the optimum data acquisition parameters of KSr 2 I 5 :Eu 2 + and compares them to NaI(Tl) for the detection of single line gamma sources between 0.01ā€“2.2 MeV with incident source strengths between 5 sāˆ’1 and 100 sāˆ’1 striking the detector. We found a strong dependence of the time-to-detection on the data acquisition parameters, where the time-to-detection of KSr 2 I 5 :Eu 2 + was 1ā€“3 times larger than NaI(Tl) across the parameter space. The largest discrepancy observed was between 0.6 MeV and 1.2 MeV, where the intrinsic 1.31 MeV beta dominates the background. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01689002
Volume :
938
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
137073090
Full Text :
https://doi.org/10.1016/j.nima.2019.05.085