Back to Search
Start Over
Rapid direct determination of tin in beverages using energy dispersive X-ray fluorescence.
- Source :
-
Talanta . Jul2019, Vol. 199, p662-666. 5p. - Publication Year :
- 2019
-
Abstract
- Abstract This article presents a new method for the simpler and faster determination of tin in beverages using EDXRF. Absorption coefficients for aqueous calibration samples were calculated and shown to be nearly identical to those of the beverage samples, thus permitting the use of aqueous standard solutions for external calibration. Beverage samples could then be measured directly using the external calibration. Determination of tin using this method takes 4 min. The LOD and LOQ were 4 mg L−1 and 15 mg L−1 respectively, and the precision was 3.89%. Different canned beverages (cold coffee, various fruit juices) were measured and the results compared to the concentrations obtained using ICP-OES after digestion. The two methods showed good compatibility, thus establishing the newly developed method as a rapid, simple, and accurate method for the determination of tin in beverages. Graphical abstract fx1 Highlights • Energy dispersive X-ray fluorescence for tin determination in beverages. • Aqueous standards and beverages absorption coefficients are nearly identical. • Beverage samples are measured directly. • Measurements are direct, simple and rapid. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00399140
- Volume :
- 199
- Database :
- Academic Search Index
- Journal :
- Talanta
- Publication Type :
- Academic Journal
- Accession number :
- 135709760
- Full Text :
- https://doi.org/10.1016/j.talanta.2019.03.024