Back to Search Start Over

Superconducting magnetic shield for MEG coupled with permalloy PC.

Authors :
Irisawa, Daiichi
Imai, Kumiko
Shintomi, Koichi
Yahara, Akihito
Matsubara, Hironori
Source :
Electrical Engineering in Japan. Apr2019, Vol. 207 Issue 1, p3-14. 12p. 9 Diagrams, 10 Graphs.
Publication Year :
2019

Abstract

The Biomagnetism measurement is required shielding effect 1/S≦1/10−7 for using SQUID magnetometer. The superconducting shield is possible to realize ideal low magnetic noise space. But it is required to attenuate every external field inside the superconducting hollow cylinder equally. Especially, radial external field decreases gently comparison to axial field. At this time the length of cylinder needs 4.4 m for diameter 1 m to realize shielding effect 1/S≦1/10−7. On the other hand, it is only 2.1 m for axial field. For that reason, we devised two methods to attenuate radial field rapidly coupled with permalloy. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
04247760
Volume :
207
Issue :
1
Database :
Academic Search Index
Journal :
Electrical Engineering in Japan
Publication Type :
Academic Journal
Accession number :
135597658
Full Text :
https://doi.org/10.1002/eej.23195