Back to Search Start Over

Robust Optical Flow Estimation in Cardiac Ultrasound Images Using a Sparse Representation.

Authors :
Ouzir, Nora
Basarab, Adrian
Lairez, Olivier
Tourneret, Jean-Yves
Source :
IEEE Transactions on Medical Imaging. Mar2019, Vol. 38 Issue 3, p741-752. 12p.
Publication Year :
2019

Abstract

This paper introduces a robust 2-D cardiac motion estimation method. The problem is formulated as an energy minimization with an optical flow-based data fidelity term and two regularization terms imposing spatial smoothness and the sparsity of the motion field in an appropriate cardiac motion dictionary. Robustness to outliers, such as imaging artefacts and anatomical motion boundaries, is introduced using robust weighting functions for the data fidelity term as well as for the spatial and sparse regularizations. The motion fields and the weights are computed jointly using an iteratively re-weighted minimization strategy. The proposed robust approach is evaluated on synthetic data and realistic simulation sequences with available ground-truth by comparing the performance with state-of-the-art algorithms. Finally, the proposed method is validated using two sequences of in vivo images. The obtained results show the interest of the proposed approach for 2-D cardiac ultrasound imaging. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
02780062
Volume :
38
Issue :
3
Database :
Academic Search Index
Journal :
IEEE Transactions on Medical Imaging
Publication Type :
Academic Journal
Accession number :
135140530
Full Text :
https://doi.org/10.1109/TMI.2018.2870947