ScatterNet: A convolutional neural network for cone‐beam CT intensity correction.

Authors :
Hansen, David C.
Landry, Guillaume
Kamp, Florian
Li, Minglun
Belka, Claus
Parodi, Katia
Kurz, Christopher
Source :
Medical Physics. Nov2018, Vol. 45 Issue 11, p4916-4926. 11p.
Publication Year :
2018

Abstract

Purpose: To demonstrate a proof‐of‐concept for fast cone‐beam CT (CBCT) intensity correction in projection space by the use of deep learning. Methods: The CBCT scans and corresponding projections were acquired from 30 prostate cancer patients. Reference shading correction was performed using a validated method (CBCTcor), which estimates scatter and other low‐frequency deviations in the measured CBCT projections on the basis of a prior CT image obtained from warping the planning CT to the CBCT. A convolutional neural network (ScatterNet) was designed, consisting of an attenuation conversion stage followed by a shading correction stage using a UNet‐like architecture. The combined network was trained in 2D, utilizing pairs of measured and corrected projections of the reference method, in order to perform shading correction in projection space before reconstruction. The number of patients used for training, testing, and evaluation was 15, 7, and 8, respectively. The reconstructed CBCTScatterNet was compared to CBCTcor in terms of mean and absolute errors (ME and MAE) for the eight evaluation patients (not included in the network training). Volumetric modulated arc photon therapy (VMAT) and intensity‐modulated proton therapy (IMPT) plans were generated on CBCTcor. Dose was recalculated on CBCTScatterNet to evaluate its dosimetric accuracy. Single‐field uniform dose proton plans were utilized for proton range comparison of CBCTScatterNet and CBCTcor. Results: The CBCTScatterNet showed no cupping artifacts and a considerably smaller MAE and ME with respect to CBCTcor than the uncorrected CBCT (on average 144 Hounsfield units (HU) vs 46 HU for MAE and 138 HU vs −3 HU for ME). The pass‐rates using a 2% dose‐difference criterion at 50% dose cut‐off, were close to 100% for the VMAT plans of all patients when comparing CBCTScatterNet to CBCTcor. For IMPT plans pass‐rates were clearly lower, ranging from 15% to 81%. Proton range differences of up to 5 mm occurred. Conclusions: Using a deep convolutional neural network for CBCT intensity correction was shown to be feasible in the pelvic region for the first time. Dose calculation accuracy on CBCTScatterNet was high for VMAT, but unsatisfactory for IMPT. With respect to the reference technique (CBCTcor), the neural network enabled a considerable increase in speed for intensity correction and might eventually allow for on‐the‐fly shading correction during CBCT acquisition. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00942405
Volume :
45
Issue :
11
Database :
Academic Search Index
Journal :
Medical Physics
Publication Type :
Academic Journal
Accession number :
132990879
Full Text :
https://doi.org/10.1002/mp.13175