Performance evaluation of the active-flow personal DataRAM PM2.5 mass monitor (Thermo Anderson pDR-1200) designed for continuous personal exposure measurements

Authors :
Chakrabarti, Bhabesh
Fine, Philip M.
Delfino, Ralph
Sioutas, Constantinos
Source :
Atmospheric Environment. Jun2004, Vol. 38 Issue 20, p3329-3340. 12p.
Publication Year :
2004

Abstract

The need for continuous personal monitoring for exposure to particulate matter has been demonstrated by recent health studies showing effects of PM exposure on time scales of less than a few hours. Filter-based methods cannot measure this short-term variation of PM levels, which can be quite significant considering human activity patterns. The goal of this study was to evaluate the active-flow personal DataRAM for PM2.5 (MIE pDR-1200; Thermo Electron Corp., Franklin, MA) designed as a wearable monitor to continuously measure particle exposure. The instrument precision was found to be good (2.1%) and significantly higher than the passive pDR configuration tested previously. A comparison to other proven continuous monitors resulted in good agreement at low relative humidities. Results at higher humidity followed predictable trends and provided a correction scheme that improved the accuracy of pDR readings. The pDR response to particle size also corresponded to previously observed and theoretical errors. The active flow feature of the pDR allows collection of the sampled particles on a back-up filter. The 24-h mass measured on this filter was found to compare very well with a Federal Reference Method for PM2.5 mass. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
13522310
Volume :
38
Issue :
20
Database :
Academic Search Index
Journal :
Atmospheric Environment
Publication Type :
Academic Journal
Accession number :
13181684
Full Text :
https://doi.org/10.1016/j.atmosenv.2004.03.007