- We describe an efficient method for determining the radii of spheres attached to atomic force microscope (AFM) cantilevers by minimizing the square root of the square error for a segment of circular arc cross-sectional data. The approach we present has benefits: (1) The method fits a function, with two fitting parameters, to a number of data points, rather than calculating the radius based on two points of cross-sectional data, (2) a fit is performed in a common spreadsheet application, using a solving feature, and (3) spheres of diameters 2 to 60 μm were measured to an accuracy of 9% and a precision of 4%. © 2004 American Institute of Physics. [ABSTRACT FROM AUTHOR]