Back to Search Start Over

Standard-deviation minimization for calibrating the radii of spheres attached to atomic force microscope cantilevers.

Authors :
Thoreson, E. J.
Burnham, N. A.
Source :
Review of Scientific Instruments. May2004, Vol. 75 Issue 5, p1359-1362. 4p.
Publication Year :
2004

Abstract

We describe an efficient method for determining the radii of spheres attached to atomic force microscope (AFM) cantilevers by minimizing the square root of the square error for a segment of circular arc cross-sectional data. The approach we present has benefits: (1) The method fits a function, with two fitting parameters, to a number of data points, rather than calculating the radius based on two points of cross-sectional data, (2) a fit is performed in a common spreadsheet application, using a solving feature, and (3) spheres of diameters 2 to 60 μm were measured to an accuracy of 9% and a precision of 4%. © 2004 American Institute of Physics. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00346748
Volume :
75
Issue :
5
Database :
Academic Search Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
13154621
Full Text :
https://doi.org/10.1063/1.1710697