A new technique for diagnosing multicharged ion beams produced by ECR ion source.
- Source :
-
Review of Scientific Instruments . May2004, Vol. 75 Issue 5, p1539-1541. 3p. - Publication Year :
- 2004
-
Abstract
- In order to study the transmission properties of multicharged ion beams between the electron cyclotron resonance (ECR) ion source and the analyzing magnet, a new diagnostic system composed of three Wien filters with three single wires has been built and installed on the IMP ECR source test bench. The single wire is used to measure the beam profile and the beam density distribution, and the Wien filter is used to measure the charge state distribution of ion beam. The single wire and the Wien filter were tested successfully. With this diagnostic system, the beam profile, charge state distributions and beam losses along the transmission line from IMP LECR3 to the analyzing magnet are experimentally studied in different conditions of the ion source. © 2004 American Institute of Physics. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00346748
- Volume :
- 75
- Issue :
- 5
- Database :
- Academic Search Index
- Journal :
- Review of Scientific Instruments
- Publication Type :
- Academic Journal
- Accession number :
- 13154573
- Full Text :
- https://doi.org/10.1063/1.1691508