Effect of Spectral Degradation and Spatio-Energy Correlation in X-Ray PCD for Imaging.

Authors :
Rajbhandary, Paurakh L.
Hsieh, Scott S.
Pelc, Norbert J.
Source :
IEEE Transactions on Medical Imaging. Aug2018, Vol. 37 Issue 8, p1910-1919. 10p.
Publication Year :
2018

Abstract

Charge sharing, scatter, and fluorescence events in a photon counting detector can result in counting of a single incident photon in multiple neighboring pixels, each at a fraction of the true energy. This causes energy distortion and correlation of data across energy bins in neighboring pixels (spatio-energy correlation), with the severity depending on the detector pixel size and detector material. If a “macro-pixel” is formed by combining the counts from multiple adjacent small pixels, it will exhibit correlations across its energy bins. Understanding these effects can be crucial for detector design and for model-based imaging applications. This paper investigates the impact of these effects in basis material and effective monoenergetic estimates using the Cramér-Rao Lower Bound. To do so, we derive a correlation model for the multi-counting events. CdTe detectors with grids of pixels with side length of $250~\mu \text{m}$ , $500~\mu \text{m}$ , and 1 mm were compared, with binning of $4\times4$ , $2\times2$ , and $1\times1$ pixels, respectively, to keep the same net 1 mm2 aperture constant. The same flux was applied to each. The mean and covariance matrix of measured photon counts were derived analytically using spatio-energy response functions precomputed from Monte Carlo simulations. Our results show that a 1 mm2 macro-pixel with $250\times 250\,\,\mu \text{m}^{\textsf {2}}$ sub-pixels shows 35% higher standard deviation than a single 1 mm2 pixel for material-specific imaging, while the penalty for effective monoenergetic imaging is <10% compared with a single 1 mm $^{\textsf {2}}$ pixel. Potential benefits of sub-pixels (higher spatial resolution and lower pulse pile-up effects) are important but were not investigated here. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
02780062
Volume :
37
Issue :
8
Database :
Academic Search Index
Journal :
IEEE Transactions on Medical Imaging
Publication Type :
Academic Journal
Accession number :
131047107
Full Text :
https://doi.org/10.1109/TMI.2018.2834369