High-pitch dual-source paranasal sinus CT in agitated patients with maxillofacial trauma: analysis of image quality, motion artifacts, and dose aspects.
- Source :
-
Acta Radiologica . Aug2018, Vol. 59 Issue 8, p909-916. 8p. - Publication Year :
- 2018
-
Abstract
- Background Image quality benefits from high-pitch scanning in agitated patients by reducing acquisition time. Purpose To compare image quality and exposure parameters in patients with maxillofacial trauma on second- and third-generation dual-source computed tomography (DSCT). Material and Methods Four groups were compared. Group 1 was examined on second-generation DSCT (120 kV/50 mAs, pitch 3.0). The other three groups were examined on third-generation DSCT. Group 2 was scanned with 120 kV/50 mAs, pitch 2.2. Automated exposure control (AEC) was used in group 3 and group 4 with pitch factors of 2.2 and 3.0, respectively. Images of third-generation DSCT were reconstructed with iterative reconstruction (IR), of second-generation DSCT with filtered back-projection. CTDIvol, acquisition time, and image quality were compared. Results Thirty patients were included in each group. Average CTDIvol (2.76 ± 0.00 mGy, 2.66 ± 0.00 mGy, 0.74 ± 0.23 mGy, and 0.75 ± 0.17 mGy) was significantly lower on third-generation DSCT with AEC ( P < 0.001). Subjective image quality was rated worst in group 4 due to strong high-pitch artifacts, while in the remaining three groups it was rated good or very good with good inter-observer agreement (k > 0.64). Average acquisition time was significantly shorter with third-generation DSCT (0.47 s, 0.36 s, 0.38 s, 0.30 s; P < 0.001). Conclusion Third-generation DSCT yields faster acquisition times and substantial dose reduction with AEC. A pitch of 2.2 should be preferred, as it results in fewer artifacts. If AEC is used, latest IR ensures that diagnostic image quality is guaranteed. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 02841851
- Volume :
- 59
- Issue :
- 8
- Database :
- Academic Search Index
- Journal :
- Acta Radiologica
- Publication Type :
- Academic Journal
- Accession number :
- 130325040
- Full Text :
- https://doi.org/10.1177/0284185117740931