Back to Search Start Over

A Bonner Sphere Spectrometer based on a large 6LiI(Eu) scintillator: Calibration in reference monoenergetic fields.

Authors :
Bedogni, R.
Pola, A.
Costa, M.
Monti, V.
Thomas, D.J.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Jul2018, Vol. 897, p18-21. 4p.
Publication Year :
2018

Abstract

A Bonner Sphere spectrometer employing a large, 11 mm diameter × 3 mm thickness, 6 LiI(Eu) scintillator (LL-BSS), was assembled. The purpose was to produce a BSS similar in sensitivity to those based on 3 He sensors, but using alternative sensors. With respect to the traditional BSS based on the 4 mm (diameter) × 4 mm (height) 6 LiI(Eu), this new BSS is a factor of 3 more sensitive. LL-BSS response matrix, determined with MCNPX, was experimentally evaluated with monoenergetic reference neutron fields of 144 keV, 565 keV and 1.2 MeV available at NPL (Teddington, UK). The results of the experiment confirmed the correctness of the response matrix within an overall uncertainty lower than ±2%. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
01689002
Volume :
897
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
129790727
Full Text :
https://doi.org/10.1016/j.nima.2018.04.040