Fully 3D geometrical calibration for an X-ray grating-based imaging system.

Authors :
Mianyi Chen
Qingsong Yang
Wenxiang Cong
Biao Wei
Ge Wang
Source :
Journal of X-Ray Science & Technology. 2016, Vol. 24 Issue 6, p821-836. 16p. 2 Color Photographs, 7 Black and White Photographs, 3 Diagrams, 2 Charts, 4 Graphs.
Publication Year :
2016

Abstract

The Precision of geometric parameters is the key requirement for grating-based phase-contrast CT and attenuationbased CT in cone-beam geometry. By extending our fan-beam geometric calibration work via Locally Linear Embedding (LLE) into the cone-beam case, here we calibrate the geometric parameters of our in-house phase-contrast CT system with attenuation projection data. Numerical and experimental studies show that the LLE-based calibration method is feasible to calibrate 8 parameters of the cone-beam geometry that improves reconstruction quality significantly. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
08953996
Volume :
24
Issue :
6
Database :
Academic Search Index
Journal :
Journal of X-Ray Science & Technology
Publication Type :
Academic Journal
Accession number :
129598996
Full Text :
https://doi.org/10.3233/XST-160590