Back to Search
Start Over
An eight-year prospective controlled study about the safety and diagnostic value of cardiac and non-cardiac 1.5-T MRI in patients with a conventional pacemaker or a conventional implantable cardioverter defibrillator.
- Source :
-
European Radiology . Jun2018, Vol. 28 Issue 6, p2406-2416. 11p. 5 Charts. - Publication Year :
- 2018
-
Abstract
- <bold>Objectives: </bold>To investigate safety and diagnostic value of 1.5-T MRI in carriers of conventional pacemaker (cPM) or conventional implantable defibrillator (cICD).<bold>Methods: </bold>We prospectively compared cPM/cICD-carriers undergoing MRI (study group, SG), excluding those device-dependent or implanted <6 weeks before enrolment or prior to 01/01/2000, with cPM/cICD-carriers undergoing chest x-ray, CT or follow-up (reference group, RG).<bold>Results: </bold>142 MRI (55 cardiac) were performed in 120 patients with cPM (n=71) or cICD (n=71). In the RG 98 measurements were performed in 95 patients with cPM (n=40) or cICD (n=58). No adverse events were observed. No MRI prolonged/interrupted. All cPM/cICD were correctly reprogrammed after MRI without malfunctions. One temporary communication failure was observed in one cPM-carrier. Immediately after MRI, 12/14 device interrogation parameters did not change significantly (clinically negligible changes of battery voltage and cICD charging time), without significant variations for SG versus RG. Three-12 months after MRI, 9/11 device interrogation parameters did not change significantly (clinically negligible changes of battery impedance/voltage). Non-significant changes of three markers of myocardial necrosis. Non-cardiac MRI: 82/87 diagnostic without artefacts; 4/87 diagnostic with artefacts; 1/87 partially diagnostic. Cardiac MRI: in cPM-carriers, 14/15 diagnostic with artefacts, 1/15 partially diagnostic; in cICD-carriers, 9/40 diagnostic with artefacts, 22 partially diagnostic, nine non-diagnostic.<bold>Conclusions: </bold>A favourable risk-benefit ratio of 1.5-T MRI in cPM/cICD carriers was reported.<bold>Key Points: </bold>• Cooperation between radiologists and cardiac electrophysiologists allowed safe 1.5-T MRI in cPM/cICD-carriers. • No adverse events for 142 MRI in 71 cPM-carriers and 71 cICD-carriers. • Ninety-nine per cent (86/87) of non-cardiac MRI in cPM/cICD-carriers were diagnostic. • All cPM-carrier cardiac MRIs had artefacts, 14 examinations diagnostic, 1 partially diagnostic. • Twenty-three per cent (9/40) of cardiac MRI in cICD-carriers were non-diagnostic. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 09387994
- Volume :
- 28
- Issue :
- 6
- Database :
- Academic Search Index
- Journal :
- European Radiology
- Publication Type :
- Academic Journal
- Accession number :
- 129492183
- Full Text :
- https://doi.org/10.1007/s00330-017-5098-z