Dipping probe electrospray ionization/mass spectrometry for direct on-site and low-invasive food analysis.
- Source :
-
Food Chemistry . Sep2018, Vol. 260, p53-60. 8p. - Publication Year :
- 2018
-
Abstract
- Rapid, direct, on-site and noninvasive food analysis is strongly needed for quality control of food. To satisfy this demand, the technique of dipping probe electrospray ionization/mass spectrometry (dPESI/MS) was developed. The sample surface was pricked with a fine acupuncture needle and a sample of ∼200 pL was captured at the needle tip. After drying the sample, the needle tip was dipped into the solvent for ∼50 ms and was moved upward. A high-voltage was applied to the needle to generate electrospray when the needle reached the highest position, and mass spectra were measured with a time-of-flight mass spectrometer. For evaluation of the method, the technique was used to analyze foods such as vegetables, salmon flesh, cow’s milk, yogurt, and soy-bean milk. The detected major ions for cow’s milk and yogurt were [(Lac) n + Ca] 2+ with n = 1–6 (where (Lac) is lactose), indicating that Ca 2+ is tightly bound by Lac molecules. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 03088146
- Volume :
- 260
- Database :
- Academic Search Index
- Journal :
- Food Chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 129273981
- Full Text :
- https://doi.org/10.1016/j.foodchem.2018.04.003