Detection of Human Immunoglobulin in Microchip and Conventional Capillary Electrophoresis with Contactless Conductivity Measurements.

Source :
Analytical Chemistry. 3/1/2004, Vol. 76 Issue 5, p1282-1288. 7p.
Publication Year :
2004

Abstract

The detection of human immunoglobulin M (IgM) was performed using capacitively coupled contactless conductivity detection (CCD) in electrophoresis carried out in conventional capillaries as well as on glass and poly(methyl methacrylate) (PMMA) microdevices. Also achieved was the analyses of IgG (an anti-human IgM) and the complex formed in the reaction between the two immunoreagents. It is demonstrated that CCD is a powerful tool suitable not only for the detection of antibodies but also for monitoring an inununological interaction. Conductivity measurements allow the direct determination of immunoreagents, and it is advantageous, since no labels are required. The immunoglobulin IgM has been taken as model analyte. The reproducibility of the analytical signal (RSD = 1%), sensitivity and limits of detection obtained for IgM (0.15 ng/mL in conventional capillaries and 34 ng/mL in microchips) are comparable to those previously obtained with amperometric detection. The immunological reaction was performed either in conventional microtiter plates as used in EUSA or in situ on the glass chip. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00032700
Volume :
76
Issue :
5
Database :
Academic Search Index
Journal :
Analytical Chemistry
Publication Type :
Academic Journal
Accession number :
12927217
Full Text :
https://doi.org/10.1021/ac0346656