Back to Search Start Over

An undulator-based spherical grating monochromator beamline for angle-resolved photoemission spectroscopy

Authors :
Hoffmann, S.V.
Søndergaard, Ch.
Schultz, Ch.
Li, Z.
Hofmann, Ph.
Source :
Nuclear Instruments & Methods in Physics Research Section A. May2004, Vol. 523 Issue 3, p441. 13p.
Publication Year :
2004

Abstract

A new spherical grating undulator beamline has been commissioned at the synchrotron radiation source ASTRID in Aarhus. It covers an energy range from 12 to <f>130 eV</f> with a resolving power of about 15.000. This beamline is combined with an end-station for angle-resolved photoemission. The spectrometer is a <f>75 mm</f> mean radius hemispherical electron energy analyzer which is mounted on a goniometer within the ultra-high vacuum chamber. The motions of the analyzer and several settings of the beamline are motorized and computerized scans can be performed. The system is in particular well suited for measurements of the electronic structure along arbitrary lines in <f>k→||</f>-space, Fermi surface mapping and studying the polarization dependence of photoemission features. [Copyright &y& Elsevier]

Details

Language :
English
ISSN :
01689002
Volume :
523
Issue :
3
Database :
Academic Search Index
Journal :
Nuclear Instruments & Methods in Physics Research Section A
Publication Type :
Academic Journal
Accession number :
12898336
Full Text :
https://doi.org/10.1016/j.nima.2004.01.039