Back to Search Start Over

An analytical X-ray CdTe detector response matrix for incomplete charge collection correction for photon energies up to 300 keV.

Authors :
Kurková, Dana
Judas, Libor
Source :
Radiation Physics & Chemistry. May2018, Vol. 146, p26-33. 8p.
Publication Year :
2018

Abstract

Gamma and X-ray energy spectra measured with semiconductor detectors suffer from various distortions, one of them being so-called “tailing” caused by an incomplete charge collection. Using the Hecht equation, a response matrix of size 321 × 321 was constructed which was used to correct the effect of incomplete charge collection. The correction matrix was constructed analytically for an arbitrary energy bin and the size of the energy bin thus defines the width of the spectral window. The correction matrix can be applied separately from other possible spectral corrections or it can be incorporated into an already existing response matrix of the detector. The correction was tested and its adjustable parameters were optimized on the line spectra of 57 Co measured with a cadmium telluride (CdTe) detector in a spectral range from 0 up to 160 keV. The best results were obtained when the values of the free path of holes were spread over a range from 0.4 to 1.0 cm and weighted by a Gauss function. The model with the optimized parameter values was then used to correct the line spectra of 152 Eu in a spectral range from 0 up to 530 keV. An improvement in the energy resolution at full width at half maximum from 2.40 % ± 0.28 % to 0.96 % ± 0.28 % was achieved at 344.27 keV. Spectra of “narrow spectrum series” beams, N120, N150, N200, N250 and N300, generated with tube voltages of 120 kV, 150 kV, 200 kV, 250 kV and 300 kV respectively, and measured with the CdTe detector, were corrected in the spectral range from 0 to 160 keV (N120 and N150) and from 0 to 530 keV (N200, N250, N300). All the measured spectra correspond both qualitatively and quantitatively to the available reference data after the correction. To obtain better correspondence between N150, N200, N250 and N300 spectra and the reference data, lower values of the free paths of holes (range from 0.16 to 0.65 cm) were used for X-ray spectra correction, which suggests energy dependence of the phenomenon. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0969806X
Volume :
146
Database :
Academic Search Index
Journal :
Radiation Physics & Chemistry
Publication Type :
Academic Journal
Accession number :
128287313
Full Text :
https://doi.org/10.1016/j.radphyschem.2018.01.003