Development of the CBM RICH readout electronics and DAQ.

Authors :
Adamczewski-Musch, J.
Akishin, P.
Becker, K.-H.
Belogurov, S.
Bendarouach, J.
Boldyreva, N.
Deveaux, C.
Dobyrn, V.
Dürr, M.
Eschke, J.
Förtsch, J.
Heep, J.
Höhne, C.
Kampert, K.-H.
Kochenda, L.
Kopfer, J.
Kravtsov, P.
Kres, I.
Lebedev, S.
Lebedeva, E.
Source :
Nuclear Instruments & Methods in Physics Research Section A. Dec2017, Vol. 876, p246-248. 3p.
Publication Year :
2017

Abstract

A real size prototype of the CBM RICH detector was tested in beam at CERN in November 2014 with new readout electronics. A detailed analysis of the timing characteristics of the readout chain will be presented in this article. Results of the time precision measurements for several subsets of all channels and the stability of the fine time calibration will be discussed. The obtained sub-nanosecond time precision allows also to investigate the effect on timing when using additional wavelength-shifting films on top of the MAPMT windows. [ABSTRACT FROM AUTHOR]