Back to Search Start Over

A performance comparison of two kinds of liquid scintillation counters from PerkinElmer, Inc.

Authors :
Feng, Xiao-gui
Jiang, Guo-hua
Huang, Jian-hua
Du, Jian-yu
He, Qian-ge
Wang, Jian-chen
Chen, Jing
Liu, Xue-gang
Source :
Journal of Radioanalytical & Nuclear Chemistry. Nov2017, Vol. 314 Issue 2, p629-635. 7p.
Publication Year :
2017

Abstract

The two kinds of liquid scintillation counters (LSCs) from PerkinElmer, Quantulus GCT 6220 and Quantulus 1220, have been compared on their basic performances, i.e., background count rate, counting efficiency, α/β discrimination, and instrument stability, by determining the radioactivity of the radionuclides of H, C, Am, Sr/Y, Co, and Cs. Except for the highlight of Quantulus GCT 6220 in stability and that of Quantulus 1220 in background for α counting, the two kinds of LSCs are comparable in the basic performances to each other. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
02365731
Volume :
314
Issue :
2
Database :
Academic Search Index
Journal :
Journal of Radioanalytical & Nuclear Chemistry
Publication Type :
Academic Journal
Accession number :
125894414
Full Text :
https://doi.org/10.1007/s10967-017-5428-2